条纹配置文件由入口配置文件的活动配置节指定,参见入口配置文件修改
条纹(检测区域)的配置节的名称由检测卡配置文件指定,参见检测卡配置文件修改
以基础版本为例,配置仅有一个用于实验模式的检测卡条纹文件内容如下:
; ====================================================================================================
; filename is selected by application
; ====================================================================================================
Author = Valiant
Copyright = cto@renshenguo.com
Linkage = http://blog.csdn.net/prsniper
Time = [2016-05-25 16:24:17]
; ====================================================================================================
; stripe configurations(select after finish parsing cassette.ini)
; ====================================================================================================
[Dummy.Stripe1]
Mode=0 ; control/test line combination
Name=[Default Stripe] ; stripe name
PosX=26.5 ; stripe X(relative to cassette, in mili-meters)
PosY=11.5 ; stripe Y(relative to cassette, in mili-meters)
SizeX=15.0 ; stripe width(relative to cassette, in mili-meters)
SizeY=4.0 ; stripe height(relative to cassette, in mili-meters)
Scale=1.0 ; stripe dilution
Rotation=0.0 ; stripe rotation(relative to cassette, in degrees)
Count=2 ; line count, max=32
0x0000=Dummy.Stripe1.Ctrl ; segment of analyte configuration in analyte.ini
0x0001=Dummy.Stripe1.Test ; line comments can make \r\n and \n no difference
配置项目如下:
Mode为检测线模式,如C/T线,Cn线,C+Tn线等
Name是条纹名称,通常用检测项目简称,如HIV,PCT,一卡多测时,通常情况是一卡一测,命名规则参见检测卡配置文件修改
PosX为条纹(检测区域)水平偏移
PosY为条纹(检测区域)垂直偏移
SizeX为条纹长度
SizeY为条纹宽度
Scale为稀释倍率,实验室应给出标准滴定的浓度和剂量,该值也可以由最终用户指定参与计算, 基础版本仅支持1.0即100%的稀释参数
Rotation为条纹的旋转角度,根据版本和仪器实际情况配置,示例为与检测卡平行
Count为检测线数量
此外是每条检测线的配置节的名称, 该节在检测线配置文件中
条纹的偏移示意图:
条纹的尺寸示意图:
单位均为毫米,参数与实际的偏差应控制在较小范围内, 以消除对检测线搜索的影响, 当然主要的影响来源于检测线的配置
但如果偏差过大, 如超出了可以搜索检测线的范围则会得到错误的结果, 通常这些参数都可以用游标卡尺精确的测量并设定