power profiler kitII 固件恢复
比如在nrf connect中意外升级了bluetooth low energy的固件导致灯不亮变砖等等
解决办法:
断开供电,卸下外壳四个螺丝钉,
将地与TP34短接(TP34在靠近USB DATA/POWER口附近)
然后上电,灯会开始闪烁
打开nrf connect中power profiler. 选择设备会提示固件要更新,更新固件后就恢复完成了
在devzone上找到的原文:
Disconnect all other nordic devices from the computer, turn off the kit and remove it from the plastic by removing the four screws. Insert USB to J2 (USB DATA/POWER). Short a connection between GND (e.g on the four pin power output header) to TP34 on the PPK2. While holding the jumper cable, turn the kit on and remove the jumper. Now the PPK2 should be in bootloader again.
Open the PPK app, and select the device which is now just a serial number.
When selecting it, the user will be prompted to program the application and it should now be working again.
链接:https://devzone.nordicsemi.com/f/nordic-q-a/90720/power-profiler-kit-ii-firmware-update-err